Descargar Imprimir esta página

Axygen AxySpin R Serie Manual De Instrucciones página 10

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 43
5.
Operation
Users can pre-set the time, temperature and speed (RPM/RCF) before a run.
Attention: Never attempt to operate the centrifuge with rotors or adapters that show signs of
corrosion or mechanical damage. Never centrifuge strongly corrosive materials that may
damage the rotors, accessories or bowl of the unit.
Setting the Time (TIME)
5.1.
Operating time can be selected from 0.5 min to 99 min or "continuous". Press
the time setting. The 2 digits of TIME on the display will flash. Turn the
increase or decrease the value. Press
(START/STOP) knob to store and run the selection. The flashing of time value will stop and the
time value will be stored automatically if no button is pressed after 6 seconds of the adjustment.
Operating time can be set as "continuous" by turning the
counter-clockwise to make the 2 digits of TIME to display "
the centrifuge will continue to run without a stop. Press
continuous operation.
Setting the Speed (RPM)
5.2.
The speed (RPM) can be selected from 500 to 13,500 rpm. Press
setting. The 5 digits of speed (RPM) on the display will flash. Turn the
increase or decrease the value. Press
(START/STOP) knob to store and run the selection. The flashing of RPM value will stop and the
RPM value will be stored automatically if no button is pressed after 6 seconds of the adjustment.
Setting the Centrifugal Force (RCF)
5.3.
The speed can be selected from 20 to 17,200 x g, too. Press
The 5 digits of RCF on the display will flash. Turn the
decrease the value. Press
knob to store and run the selection. The flashing of RCF value will stop and the RCF value will be
stored automatically if no button is pressed after 6 seconds of the adjustment.
Setting the Temperature (TEMP)
5.4.
The temperature can be selected from -10 to 40°C. Press
setting. The 2 digits of temperature on the display will flash. Turn the
increase or decrease the value. Press
(START/STOP) knob to store and run the selection. The flashing of TEMP value will stop and the
value will be stored automatically if no button is pressed after 6 seconds of the adjustment.
AxySpin R™ Refrigerated Microcentrifuge
7
button again to confirm the selection or press
button again to confirm the selection or press
button again to confirm the selection or press
button again to confirm the selection or press
© 2016 Corning Incorporated. All Rights Reserved.
(START/STOP) knob to
(START/STOP) knob clockwise or
- -
". When the time is set as "Continuous",
(START/STOP) knob to end the
button to change the speed
(START/STOP) knob to
button to change the RCF setting.
(START/STOP) knob to increase or
button to change the temperature
(START/STOP) knob to
button to change
(START/STOP)