Cooling
5.5.
The temperature can be set from -10 °C and + 40 °C. The set temperature can also be changed
during the run.
Pre-Cooling
Pre-Cooling starts a temperature-control run at 6,000RPM. It allows the rotor to be cooled
down to the new, pre-set temperature quickly. Cooling down from room temperature to 4
°C takes approximately 8 min. (max. 16 minutes.) To activate Pre-Cooling function, press
and hold
button, then press
will flash during Pre-Cooling. The Pre-Cooling can be stopped by pressing the
(START/STOP) key. The Pre-Cooling will function only when the set temperature is
lower than the ambient temperature.
Standby Cool
ng
i
When the centrifuge is powered on, the refrigeration system will cool down the chamber to
the pre-set temperature before or after the run if the chamber temperature is above the
pre-set temperature.
Note:
1. To extend the life cycle of the compressor, always remember to inactivate the Standby
Cooling (by increasing the set temperature above room temperature) or power off the
centrifuge when it is not in use.
2. When the lid is open, the Standby Cooling will stop after 3 minutes.
Quick-Spin Operation
5.6.
The centrifuge can be operated for a short run by pressing and holding the
will continue to run as long as the
the time display. The Quick-Spin speed and temperature is based on the last run settings remaining in
the memory. The icon "RPM" will flash during Quick-Spin operation.
Changing the Settings During a Run
5.7.
Users can change the settings while the rotor is running. Through a one-time pressing of any function
buttons (
,
,
The to-be-changed value flashes, and can then be changed. Once the display switches into the
current value mode after completion of the entered values, the new values are activated.
The new values are stored and executed by pressing the function button again or by waiting 6
seconds.
6.
Service and Maintenance
Always unplug the power plug before service and maintenance.
8
(START/STOP) key. Both "RPM" and "TEMP" icons
button is depressed and the time, in seconds, will count up on
,
) , the current value will switch into the pre-set value mode.
© 2016 Corning Incorporated. All Rights Reserved.
button. The centrifuge