Initial Operation
3.2.
Place the machine on a stable, flat table. Please keep the machine at least 30 cm from the
surrounding area or objects.
Mains power and frequency "MUST" be compatible with the information given on the product label on
the rear of the centrifuge.
Connect the power cord to the power socket of the centrifuge, then to the power source. Switch the
Power On/Off switch to turn on the centrifuge. After the LCD display is on and the buzzer beeps twice,
the centrifuge is ready to operate.
Opening and Closing the Lid
3.3.
The lid can be opened only when the centrifuge is powered on. Press
The lid will be locked automatically when the user presses down the front edge of the lid.
Note:
Do not slam the lid!
Loading the Rotor
3.4.
For safety reasons, the rotor must be loaded with centrifuge tubes symmetrically. The difference in
the weight between the tubes should not exceed 0.1 gram.
The centrifuge tubes should be loaded in pairs opposite each other and each should contain roughly
the equal weight of sample. One or two additional loaded tubes may need to be added to achieve this.
Please refer Figure 1 to see an example of a properly loaded rotor.
Attention: To reduce air friction and noise, the rotor lid should be attached to the rotor during
centrifugation. Make sure that the rotor lid snaps securely into place, by pressing down on
center catch.
AxySpin R™ Refrigerated Microcentrifuge
3
© 2016 Corning Incorporated. All Rights Reserved.
button to open the lid.