User Manual
3.4
Scope of delivery
Component
Wall-mounted electronics and control unit
Consisting of the keypad and display, power, signal and Modbus connections (if equipped)
Ultrasonic Flow Sensors
Two transducers for flow measurement with mounting fixture and cables
Guide rail
Gel-Pads
Quick release pipe clamps
Modbus Cable (Modbus version only)
PT100 temperature sensor with 3 meter cable (HM version only)
Stainless cable tie for PT100 temperature sensors (HM version only)
12 V DC power supply and adapters (optionally supplied)
Manual
4
Installation
4.1
Identify suitable location for flow measurement
The GF U1000 V2 WM/WHM requires an even and uniform flow profile, since distortions in the flow can cause unpredictable
measurement errors.
In many applications, however, it is not possible to have a uniform flow rate through 360°. This may be, for example, because there
are air bubbles inside the top of the pipe, or turbulence in the pipe, or sludge at the bottom of the pipe.
Possible air gap
Possible deposits
(sludge)
5x inner diameter
Experience has shown that the most accurate results are obtained if the guide rail of the transducers is not mounted vertically to
the pipe, but turned at an angle of about 45° to the right
or left on the pipe.
Turbulent flow profile
Possible sensor position
Laminar flow profile
10x inner diameter
Installation
Pcs.
1
1
1
4
2
1
2
4
1
1
Flow
11