Safety
1.4.2
Name plate
1.4.3
Name plate: TTF300 - HART
The name plate is located on the transmitter housing.
Fig. 1: Example for ATEX explosion protection
1 Project number
2 Manufacturer of transmitter
3 Country and year of manufacture
4 Order number
5 Serial number
6 Supply voltage range, typical current
range, protocol
7 Customer configuration
8 Ambient temperature range / housing
degree of protection
6 - EN
1
2
Automation Products GmbH
Alzenau
3
4
O-Code: TTF300-E1H
5
Ser.-No: 2065342
6
U = + 11...30 V, I = 4...20 mA, HART
S
a
7
Cfg: 1 x Pt100 3W; 0...100°C
8
TA. = -40°C...+85°C
TTF300-E1H;
9
II 1 G Ex ia IIC T6
II 2(1) G Ex [ia]ib IIC T6
II 2G (1D) Ex [iaD]ib IIC T6
T1...T4 Ta. = -40 °C ... +60 °C (Zone 0) ... +85 °C (Zone 1)
T5
Ta. =
-40 °C ... +56 °C (Zone 0) ... +71 °C (Zone 1)
T6
Ta. =
-40 °C ... +44 °C (Zone 0) ... +56 °C (Zone 1)
10
11
Germany 2008
IP20
PTB 05 ATEX 2017 X
9 Ex mark (optional)
10Protection class of hazardous area
design (optional)
11Temperature class of hazardous area
design (optional)
12Refer to product documentation
13Software revision number / hardware
revision number
14CE mark (EC conformity)
15Type designation
16Safety integrity level (optional)
TTF300
TTF300
2400045678
HW. Rev. 1.06
SW. Rev. 1.01.03
TTF300
16
15
14
13
12
A00265
CI/TTF300-X1