cycle At 0°, 45°,
90°, 135°, 180°,
225°, 270°and
Voltage dips,
short interrup-
tions and voltage
0 % UT; 1 cycle
variations on
power supply
70 % UT; 25/30
input lines
IEC 61000-4-11
0 % UT; 250/300
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
The device is intended for use in the electromagnetic environment specified below. The customer or the user of
device should assure that it is used in such an environment.
Immunity
IEC 60601
test
test level
3V
0.15 MHz to
Conducted
80MHz
RF
6 V in ISM
IEC 61000-
and amateur
4-6
radio bands
between
0.15 MHz
and 80 MHz
Radiated RF
10V/m
IEC 61000-
80 MHz a
4-3
2,7 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp-
tion and reflection from structures, objects and people.
0 % UT; 0.5
315°
and
Not applicable
cycles
Single phase:
at 0°
cycles
30 A/m
Declaration - electromagnetic immunity
Compliance
level
Not
applicable
10V/m
Mains power quality should be that of a
typical commercial or hospital environ-
ment. If the user of the device requires
continued operation during power mains
interruptions, it is recommended that the
device be powered from an uninterrupti-
ble power supply or a battery.
Power frequency magnetic fields should
be at levels characteristic of a typical lo-
30 A/m
cation in a typical commercial or hospital
environment.
Electromagnetic
environment - guidance
Portable and mobile RF communications equipment should be
used no closer to any part of device, than the recommended
separation distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
150KHz to 80 MHz
80MHz to 800 MHz
80MHz to 2,7 GHz
where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, a should be less than the compliance
level in each frequency range.
Interference may occur in the vicinity of equipment marked with
the following symbol: