Disinfecting the test strip guide cover and test strip guide
NOTE: Ensure the meter surface stays
moist for the entire contact time.
3
Dry the meter thoroughly with a dry cloth
or gauze. Visually verify that no solution
is seen anywhere on the meter at the
completion of cleaning and disinfecting.
Ensure that the meter is thoroughly dried
after cleaning and disinfecting.
NOTE: Clean the test strip guide prior to each
disinfection step.
CoaguChek Vantus System User's Manual, version 2.0