CoaguChek Vantus System User's Manual, version 2.0
NOTICE
Damage to the instrument
• Ensure that no liquid enters the meter
when disinfecting the test strip guide. If
moisture enters the meter, it may cause
malfunction of the instrument.
• Do not insert any objects in the test strip
guide. Doing so may damage the electrical
contacts behind the test strip guide.
• Remember to keep the meter on a level
surface prior to disinfecting and powered
off. Squeeze off excess disinfecting solution
or blot on a dry paper towel to remove
any excess disinfecting solution before
disinfecting the surface of the meter.