144
Error
Meter Lifetime End
[E-33]
Maximum number of tests has
been reached. Contact your
healthcare professional or
Roche Diagnostics.
19 Jul 2018
10:10 AM
Error
Meter Error [E-62]
Temperature or heater error.
Power meter off and on. If
error persists, contact Roche
Diagnostics.
19 Jul 2018
10:10 AM
Error
Meter Error [E-51]
Meter self-check detected a
general error. Power meter off
and on. If error persists,
contact Roche Diagnostics.
19 Jul 2018
10:10 AM
Error
Meter Error [E-63]
Hardware or electronic error.
Power meter off and on. If
error persists, contact Roche
Diagnostics.
11 Jun 2018
10:10 AM
These error messages may appear on the
display. They have the meanings shown:
Error
Meter Error [E-52]
Temperature or heater error.
Please try again in 3:15
minutes. If error persists,
contact Roche Diagnostics.
19 Jul 2018
10:10 AM
Error
Meter Error [E-64]
Test strip reader error. Power
meter off and on. If error
persists, contact Roche
Diagnostics.
11 Jun 2018
10:10 AM
CoaguChek Vantus System User's Manual, version 2.0
Error
Meter Error [E-61]
Meter is defective. No test is
possible. Contact Roche
Diagnostics.
19 Jul 2018
10:10 AM
Error
Meter Error [E-65]
Code chip cannot be read.
Replace with a different code
chip.
11 Jun 2018
10:10 AM