Accuracy Table
Final test values related to the nominal capacity (maxi-
mum volume) or the indicated volume steps indicated on
the instrument, obtained when instrument and distilled
water are equilibrated at ambient temperature (20 °C/
68 °F) and with smooth operation. According to DIN
EN ISO 8655.
Transferpette
Volume range
µl
0.5 - 10
5 - 50
10 - 100
20 - 200
30 - 300
* A = Accuracy
* CV = Coefficient of Variation
38
S -8/-12
®
Volume step
µl
10
5
1
50
25
5
100
50
10
200
100
20
300
150
30
A*
CV*
Increment
≤ ± %
≤ %
µl
1.6
1.0
0.01
2
2
8
6
0.8
0.4
0.1
1.4
0.8
6
3
0.8
0.3
0.2
1.4
0.6
4
2
0.8
0.3
0.2
1.4
0.6
4
1.5
0.6
0.3
0.5
1.2
0.6
3
1.5
20 °C
!
Ex
Rec. type
of tips, µl
0.5 - 20
2 - 200
2 - 200
2 - 200
5 - 300