Disassembling the Manifold
➃
➄
➅
44
4.
Push the piston unit down to the
bottom. The seal will either remain
inside the nose cone or will stay on
the piston after the nose cone is
removed. Remove the seal, inspect it
and clean or replace if necessary. If
required, lubricate the piston lightly
with the supplied silicone oil.
(In the Transferpette
30-300 µl unit, remove the additional
pressure ring from the piston for
cleaning.)
5.
Push the seal on the piston with its
flat side facing upward.
(In the Transferpette
30-300 µl unit, first replace the
additional pressure ring!)
Use the mounting tool to mount and
tighten the cleaned or new nose cone.
6.
Reassemble the manifold. The mani-
fold must be screwed into the handle
counter-clockwise, until it audibly
snaps into place.
Check the instrument for tightness
(see page 42, "Servicing").
S -8/-12
®
S -8/-12
®