Descargar Imprimir esta página

Telstar Bio II Advance Plus Serie Manual De Usuario página 11

Cabinas de seguridad microbiológica

Publicidad

Idiomas disponibles

Idiomas disponibles

Fig. 1
WARNING
For any handling during assembly or dismantling of the cabinet, appropriate safety measures
should be taken, such as wearing safety gloves or shoes.
c) Cabinets should be placed away from entrances, corridors, ventilation supply points, etc.,
which might affect airflow in the work area.
d) Ensure the cabinet is set perfectly on the table or workbench in order to prevent
vibrations.
e) To prevent the work area from overheating, we recommend that the room temperature
is always within the comfort margins for working.
The optimal comfort margins are:
Temperature: 18-28ºC
-
Relative humidity: 50%±20%
-
If any of these conditions is not fulfilled, optimum cabinet performance cannot be
guaranteed.
f)
When necessary, air vents in the room should be modified or speed of air entry reduced
so that air is not aimed directly at the cabinet, as this may produce turbulence.
Side air vents are not appropriate for rooms where cabinets are installed.
UM-BioIIAdvancePlus-EN-1812
X
Fig. 3
Fig. 2
9 / 35

Publicidad

loading