Telstar Bio II Advance Plus Serie Manual De Usuario página 21

Cabinas de seguridad microbiológica

Publicidad

Idiomas disponibles

Idiomas disponibles

4.5.
CONTROL KEYBOARD-DISPLAY
The following elements are found in the control keyboard:
A. On/Off Button
Pressing when the cabinet is turned off:
It goes to the Password Request screen. When entering the correct password (Chapter
4.5 point 1), it goes to the Main Operation screen.
Pressing when the cabinet is turned on:
It goes to the Password Request screen. When entering the correct password (Chapter
4.5 point 1), the cabinet switches off.
B. Navigation buttons
They move the cursor through all different function icons and select them. The icon
pointed inverts its colours. If the selected item is a numeric field, the navigation buttons
increase/decrease and validate the value.
C. Backlight graphic display (Annex 3 pos. 3):
It shows alarm messages, operating graphs, function icons and all the information
required in cabinet-user communication. The background colour changes to:
Green: Working mode. All the flows are working fine and there are no alarms
-
activated.
Red: Alarm mode. Any alarm has been activated or the cabinet is in unsafe mode. In
-
case of alarm, an acoustic alarm is activated together with the red light.
Yellow: Disinfection mode.
-
White-blue: Information mode. It is used in Technical Service screens.
-
Below is an explanation of the different system screens:
1. Password Request screen (by default 0000)
UM-BioIIAdvancePlus-EN-1812
(Annex 3 pos. 1):
,
and
(Annex 3 pos. 2):
Right
-
Wrong Password: Cursor returns to the first
-
Increase the value of the selected field.
Decrease the value of the selected field.
Confirm the value of the selected field and
move the selection to the next field. After
pressing for the final time:
Password:
it
accesses
Operation screen.
field.
the
Main
19 / 35