q) People traffic generates air movements that may cause unexpected loss of the
containment barrier and, in the case of Class II cabinets, a reduced level of product
protection. Warning signs and other means should be used to avoid traffic during critical
operations.
Cabinet locations can be divided to reduce the effects of air movement produced by other
cabinets and people traffic.
r)
Suddenly doors opening and closing may disturb airflows in the cabinet and it should be
avoided when cabinets are being used.
WARNING
The UV light system can only be used when the front glass is closed.
Never connect equipment whose consumption exceeds the maximum one indicated on the
internal sockets of the cabinets.
4.3.1.
ERGONOMICS
Microbiological security cabinet Bio II Advance Plus and its accessories have been designed
following recommendations and specifications of the standard EN ISO 14738:2010.
Users should consider their position when working with the cabinet, for their own safety.
Below are examples of four positions NOT recommended and one that IS recommended.
Fig. 1
UM-BioIIAdvancePlus-EN-1812
X
X