3.3 Temperature Calibration

1.
TO START
2.
PREPARATION
TARE
3.
START CALIBRATION
100C
TARE
4.
AFTER 15 MINUTE TIME PERIOD
NOTE: Optional Temperature Calibration Kit is required.
ANALYZER SETUP
ANALYZER SETUP
ANALYZER SETUP
ANALYZER SETUP
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
TEMP CAL
TEMP CAL
TEMP CAL
TEMP CAL
TEMP CAL
REMOVE PAN SUPPORT
ABORT CALIBRATION
TEMP CAL 100C
TEMP CAL 100C
TEMP CAL 100C
TEMP CAL 100C
TEMP CAL 100C
CURRENT TEMP:
--- C
TM TO CAL PNT:
15:00 MIN
ADJ CAL READING:
100 C
ACCEPT NEW CAL
ABORT CALIBRATION
TEMP CAL 100C
TEMP CAL 100C
TEMP CAL 100C
TEMP CAL 100C
TEMP CAL 100C
CURRENT TEMP:
100 C
TM TO CAL PNT:
10:00 MIN
ADJ CAL READING:
101 C
ACCEPT NEW CAL
ABORT CALIBRATION
x1
15 minute time lapse
or
ADJUST TO THERMOMETER READING
TARE
EN-9
ANALYZER SETUP
ANALYZER SETUP
ANALYZER SETUP
ANALYZER SETUP
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
TEMP CAL
TEMP CAL
TEMP CAL
TEMP CAL
TEMP CAL
INSERT CAL FIXTURE
CLOSE COVER
ABORT CALIBRATION
TEMP CAL
TEMP CAL
TEMP CAL
TEMP CAL
TEMP CAL
CALIBRATION READY
PRESS TARE/ACCEPT
ABORT CALIBRATION
TEMP CAL 100C
TEMP CAL 100C
TEMP CAL 100C
TEMP CAL 100C
TEMP CAL 100C
CURRENT TEMP:
100 C
TM TO CAL PNT:
10:00 MIN
ADJ CAL READING:
100 C
ACCEPT NEW CAL
ABORT CALIBRATION
TEMP CAL 100C
TEMP CAL 100C
TEMP CAL 100C
TEMP CAL 100C
TEMP CAL 100C
CURRENT TEMP:
100 C
TM TO CAL PNT:
09:23 MIN
ADJ CAL READING:
101 C
ACCEPT NEW CAL
ABORT CALIBRATION