Descargar Imprimir esta página

Medline Hemo-Force MDS600INT2 Instrucciones De Operación página 17

Bomba para tvp intermitente 2nda generación
Ocultar thumbs Ver también para Hemo-Force MDS600INT2:

Publicidad

Idiomas disponibles

Idiomas disponibles

Technical Data
Model No.
Item Name
Ref No.
Size
Weight
Pressure Range
Input Rating
Power Cable
Fuse Rating
Classification
Ingress Protection
Operation Humidity
Operation Temperature
Operation Atmospheric
Pressure Range
Mode of Operation
Applied Part
Battery Pack Model.
(optional)
Hemo-Force®
Intermittent DVT Pump
MDS600INT2
7.54"x 5.12"x7.95"
5 lbs (2.8 Kg) - without battery
Calf/Thigh Garment: 40 mmHg +10/-5mmHg (5.3kPa
+1.3kPa/-0.7 kPa)
Foot Garment: 120 mmHg +10/-5 mmHg (16 kPa +1.3/-
0.7 kPa)
AC 100-240V, 50/60Hz, 0.6~0.4A
15ft, non-shielding, AC powered (USA)
T1AH 250V
Class II, Type BF
Not AP or APG type
IP22
30–75%
15–35°C
700 hPa to 1060 hPa
Continuous
Garment
MDS600BAT (14.4 Vdc, 2200 mAh / 31.68 Wh
Lithium-ion rechargeable battery pack)