Medline Hemo-Force MDS600INT2 Instrucciones De Operación página 21

Bomba para tvp intermitente 2nda generación
Ocultar thumbs Ver también para Hemo-Force MDS600INT2:

Publicidad

Idiomas disponibles

Idiomas disponibles

Manufacturer's Declaration – Electromagnetic Immunity
The Hemo-Force® is intended for use in the electromagnetic environment (for professional
healthcare) specified below. The customer or the user of the Hemo-Force® should assure that it
is used in such an environment.
Immunity test
IEC 60601 test level Compliance level
Conducted RF
3 Vrms:
IEC 61000-4-6
0,15 MHz – 80 MHz
6 Vrms:
in ISM bands
between
0,15 MHz and 80
MHz
80 % AM at 1 kHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms:
in ISM bands
between
0,15 MHz and 80
MHz
80 % AM at 1 kHz
3 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
21
Electromagnetic
environment-guidance (for
professional healthcare
environment)
Portable and mobile RF
communications equipment
should be used no closer to
any part of the Hemo-Force®
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance:
d = 1,2
d = 1,2
80MHz to 800 MHz
d = 2,3
800MHz to 2,7 GHz
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Interference may occur in the
vicinity of equipment marked
with the following symbol: