Manufacturer's Declaration – Electromagnetic Immunity
The Hemo-Force® is intended for use in the electromagnetic environment (for professional
healthcare) specified below. The customer or the user of the Hemo-Force® should assure that it is
used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic
Contact: ±8 kV
discharge (ESD)
Air±2 kV, ±4 kV, ±8
IEC 61000-4-2
kV, ±15 kV
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/
output lines
Surge
± 0.5 kV, ±1 kV
IEC 61000-4-5
line(s) to line(s)
± 0.5 kV, ±1 kV,± 2kV
line(s) to earth
Voltage Dips,
Voltage dips:
short interrup-
0 % UT; 0,5 cycle
tions and voltage
0 % UT; 1 cycle
variations on
70 % UT; 25/30
power supply
cycles
input lines
IEC 61000-4-11
Voltage
interruptions:
0 % UT; 250/300
cycle
Power frequency
30 A/m
(50, 60 Hz)
50 Hz or 60 Hz
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Compliance level
Contact: ±8 kV
Air±2 kV, ±4 kV, ±8
kV, ±15 kV
± 2kV for power
supply lines
Not applicable
± 0.5kV, ± 1kV line(s)
to line(s)
Not applicable
Voltage dips:
0 % UT; 0,5 cycle
0 % UT; 1 cycle
70 % UT; 25/30
cycles
Voltage
interruptions:
0 % UT; 250/300
cycle
30 A/m
50 Hz and 60 Hz
20
Electromagnetic
environment-guidance (for
professional healthcare
environment)
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Mains power quality
should be that of a typical
professional healthcare
environment.
Mains power quality
should be that of a typical
professional healthcare
environment.
Mains power quality
should be that of a typical
professional healthcare
environment. If the user of
the Hemo-Force® requires
continued operation during
power mains interruptions,
it is recommended that the
Hemo-Force® be powered
from an uninterruptible
power supply or a battery.
The Hemo-Force® power
frequency magnetic fields
should be at levels
characteristic of a
typical location in a typical
professional healthcare
environment.