revPIP Mode
Program for pipetting of liquids with high viscosity, vapor pressure or that tend to foam.
Volume and speed adjustments are described on pages 40 and 41.
What to do
1.
Bring up the menu
2.
Select revPIP
mode
3.
Confirm revPIP
mode
4.
Prepare for
pipetting
5.
Aspirate liquid
6.
Discharge liquid
in the revPIP
mode
4 8
How to do it
Press the MENU key three
times to bring up the mode
selection menu.
'Mode' blinks.
Scroll through the modes
using the arrow keys until
'revPIP' appears.
'Mode' continues to blink.
Press the ENTER key.
The Display now shows
'blo' for blow-out.
Press the pipetting key
once to move the pistons
into the start position.
The arrow in the display
points upwards (aspiration).
Press the pipetting key
once. The volume
aspirated will be a little
bit more than set.
To discharge the measured
amount of liquid, press the
pipetting key once. The
arrow in the display points
downwards (discharge).
Some liquid will remain
in the tips.
The Programs · Settings and Use
Keys to press
Display readout
3x
1x
1x
1x