6. Priming the instrument
a.
b.
30 mm
997468
a.
Hold the discharge tube and remove the closure cap/
screw cap.
b.
To avoid splashes, hold discharge tube orifice on the in-
ner wall of a suitable receiving vessel.
For priming, let the piston move up approx. 30 mm and
push it down rapidly until the lower stop.
Repeat this procedure approximately 5 times until there
are no more air bubbles under the piston. Bubbles less
than 1 mm are permitted.
Before using the instrument for the first time, ensure it is
rinsed carefully and discard the first few samples dispensed.
Avoid splattering.
NOTICE
Operating manual
4 Assembly
59