10. Technical specifications
Test voltages
Maximum resistance reading
Short circuit current
Test voltages accuracy
Equipment basic accuracy
Advanced features
Built-in printer (optional)
Interface
Memory up to 4000 measured
values
Built-in chronometer
Real time clock
Environmental protection
Safety class
EMC
Electromagnetic irradiation
Immunity
Electrostatic immunity
: 500, 1000, 2500, 5000 V with fast selection.
500 V to 5 kV in 100 V or 500 V steps.
DC, negative in relation to grounding.
: 5 T @ 5 kV.
: 1.5
0.5 mA
:
3% of nominal value over 10 G resistance.
:
5% of reading
3 digits
(1 M to 500 G at any test voltage).
: Automated Polarization Index computing
Automated Dielectric Absorption Ratio
computing
Step Voltage Test
Programmable timer.
"Pass-fail test" with programmable limits.
: Prints elapsed time, actual voltage applied to
the charge and measured resistance.
: USB.
: It allows for the storage of 4000 tests readings
in its internal NVRAM memory.
: Shows elapsed time since measurement stats
in mm:ss format.
: Indicates date, hour and minutes.
: IP54 (with closed lid).
: In accordance with IEC 61010-1.
: In accordance with IEC 61326-1
: In accordance with IEC 61000-4-3
: In accordance with IEC 61000-4-2
26