07
Información del producto
Recommended protective distances between portable
and mobile HF telecommunication devices and custo screen 200
custo screen 200 is designed for the operation in an electromagnetic environment in which the HF
transients can be controlled. The user can help avoid electromagnetic interferences by keeping to the
minimum distance between portable and mobile HF telecommunication devices (transmitters) and the
device – depending on the initial performance of the communication device, as indicated below.
Nominal power
of the transmitter
W
0,01
0,1
1
10
100
For transmitters whose maximum nominal power is not indicated in the above table, the recommended
protective distance d can be determined in meters (m), using the equation affi liated with the correspon-
ding column. P is the maximum nominal power of the transmitter in Watt (W) according to the indications
of the manufacturer of the transmitter.
COMMENT 1:
With 80 MHz and 800 MHz the higher frequency range is valid.
COMMENT 2:
These guidelines may not apply in every case. The propagation of electromagnetic factors is infl uenced by
absorptions and refl ections of buildings, objects and people.
Table 206 according to DIN EN 60601-1-2, 6.8.3.201
40
Protective distance depending on the transmitting frequency in m
150 kHz to 80 MHz
d= (3,5/V1) √P
non-applicable
non-applicable
non-applicable
non-applicable
non-applicable
Monitorización presión arterial con custo screen 200 y custo diagnostic | GEB 0116 – DK 0897 | Versión 001 – 22.05.2009 | custo med GmbH
80 MHz to 800 MHz
800 MHz to 2,5 GHz
d= 1,2 √P
0,12
0,38
1,20
3,79
12,00
d= 1,2 √P
0,23
0,73
2,30
7,27
23,00