Adjustment range
Nominal volume [ml]
1
2
5
10
25
50
100
8.2 Analog-adjustable model
a.
b.
Adjustment range
Nominal volume
1 ml
2 ml
5 ml
10 ml
25 ml
50 ml
100 ml
997490
Digital max. +/- [µl]
12
24
60
120
300
600
–
a.
Insert the pin of the assembly tool into the cover plate
(Pos. 1), and break it off with a rotating motion. Discard
1
the adjustment cover.
b.
Insert the pin of the assembly tool into the adjustment
screw and rotate to the left in order to increase the dis-
pensing volume, or rotate to the right to decrease the
dispensing volume (e.g. for an actual value of 9.97 ml,
rotate approx. 1/2 turn to the left).
c.
Adjustment is completed.
ð
The change in the adjustment is indicated by a red disk
(circle in figure).
Analog-adjustable/Fixed-
volume max. +/- [µl]
6
12
30
60
150
300
600
8 Adjustment
One rotation corresponds
to [µl]
~ 8
~ 16
~ 40
~ 80
~ 130
~ 265
~ 400
Operating manual