Service & maintenance information
Service and maintenance of a medical device is wholly the responsibility of the owner of that device.
Failure to service and maintain a device in accordance with the instructions may invalidate the
device's warranty. Furthermore, failure to service and maintain a device may compromise the clinical
condition or safety of users and/or their caregivers. Do not perform service and maintenance while the
user is in the product. Contact the local dealer for assistance, if needed, in setting up, using or how to
perform service and maintenance of the device.
Lifetime
The lifetime of this product in normal use
is 8 years if all maintenance and servicing
is carried out in accordance with the
manufacturer's instructions and demonstrably
recorded. The device may be reissued within
its lifetime.
Service interval
The frequency of inspection may be
altered depending upon use and wear. It is
recommended that the product is inspected
yearly, every time the product is re-issued for
use and after a long-term storage (more than
4 months). The inspection should be carried
out by a person who understands the use of
the product.
Service checklist:
Check and adjust the following as per their
intended use:
• For products with gasspring and/or
telescopic tube, check that the locking
mechanisms work. Also check that it is
possible to tilt the basis plate horizontally
and vertically.
• For products with power, recharge the
battery and check that the handset works
by tilting the basis plate to vertical and
horizontal position
• Check that all castors are moving freely
and that all wheel locks/brakes are easily
activated
• The foot plate(s) is adjustable, both
distance to basisplate and the angle
mechanism must be checked
• All cushions are in good shape without
cracks and or holes in the cover
• Nuts and bolts on the product are
tightened
• The straps must be without cracks and not
flossed, pay special attention to the area
where the straps connect to the brackets.
• The plastic pieces connecting the frame to
the basis plate is undamaged and without
cracks. Pay special attention to the areas
where fasteners are inserted
• The rails in the basis plate must be in good
shape, it should be possible to slide a
strap up and down the rail.
• All labels on the product are intact
• Check that all cables connecting the power
system are in good shape.
Repair or replace any damaged or worn out
EN
parts.
Maintenance
Before every use it is recommended to:
Wipe off any residue and dirt from the product,
using a cloth with warm water and a mild
detergent/soap without chlorine and let it dry
before use carry out a visual inspection, for
any damaged or worn-out parts.
Washing
Main product
This product can be washed at 60° using a
mild detergent for 10 minutes in a washing
machine designed for medical devices. Use
the machine's drying function to dry the
product. Unmount all accessories and wash
them separately.
Before washing
Avoid degreasing of the piston rods, the
actuators should be retracted to minimum
stroke before washing.
• Remove all accessories, unplug all cables
and make sure to dismount the battery,
control-box and handset. Leave the
actuators with the cables mounted on the
product
• Do NOT clean the power unit with a high
pressure cleaner or steam cleaner.
After washing
• Ensure correct placement of cables
inserted into the control box when
reinstalling the power unit system
• Check that the handset functionalities
works correct
Handwash
This product and its accessories can also be
washed by hand. Use warm water and a mild
detergent/ soap without chlorine and let all
parts dry before use.
Disinfection
The product can be disinfected with a 70%
disinfectant IPA solution. It is recommended to
wipe off any residue and dirt from the product,
using a cloth with warm water and a mild
detergent/soap without chlorine and let it dry
before disinfection.
Materials
• Stainless steel
• Aluminium
• Plastic
28