Multimeter Features
3
– Changing the range manually will also restart the recording session.
NOTE
– If an overload is recorded, the averaging function will be stopped.
is shown in place
of the average value.
– The APO (Auto Power-Off) function is disabled when MaxMin is enabled.
– The maximum recording time is 99999 seconds (27 hours, 46 minutes, 39 seconds).
is shown if the recording exceeds the maximum time.
4 Press
for more than 1 second to disable the MaxMin function.
MaxMin
This mode is useful for capturing intermittent readings, recording minimum and maximum
readings unattended, or recording readings while equipment operation keeps you from
observing the multimeter display.
The average value displayed is the true arithmetic mean of all readings taken since the
start of recording. The average reading is useful for smoothing out unstable inputs,
calculating power consumption, or estimating the percentage of time a circuit is active.