Alerts-Continued
Alarm item
Battery
Exhausted
Replace
Sieve Bed
Low Input
Voltage
Absence
Of Breath
Oxygen
Concentration <87%
Low Battery
Oxygen
Concentration
<50%
Breath Sensor
Fail
Oxygen Sensor
Fail
Gas Delivery Fail
Gas Obstruction
Tank Pressure
Fail
Alarm condition
Battery cycle >500
Or battery health <50%
Sieve Bed expired or
Sieve Bed chip error
Adapter input
<17.0v
No breath detected
continuously for more
than 15 seconds
Concentration <87%
continuously for more
than 300 seconds
5%
RSOC
20%
Without adapter
Concentration <50%
continuously for more
than 300 seconds
Breath Sensor failed
Oxygen Sensor failed
No delivery detected
after injection
Pipe or nasal blocked
Tank pressure failed
System process
Alarm only
Alarm only
Auto-switch to battery
until the
adapter input>18v
Alarm only
Alarm only
Alarm only
Auto-shut down
after 30 seconds
Auto-shut down
after 30 seconds
Auto-shut down
after 30 seconds
Auto-shut down
after 30 seconds
Auto-shut down
after 30 seconds
Auto-shut down
after 30 seconds
14
Display of screen
Battery is Exhausted.
Replace Battery
Replace Sieve Bed
Contact Provider
Low Input Voltage
Check Adapter
Absence of Breath
Check Cannula
Low O2: <87%
Contact Provider
Low Battery
Charge Now
Low O2: <50%
Contact Provider
Breath Sensor Fail
Contact Provider
Oxygen Sensor Fail
Contact Provider
Gas Delivery Fail
Contact Provider
Gas Obstruction
Contact Provider
Tank Pressure Fail
Check Provider