Operating Instructions
General Operation
1. Find a well ventilated location to place the P2-E6. Ensure that the unit is turned
off.
Ensure the Intake and exhaust has clear access. The P2-E6 should be placed where any
alarms can be heard.
WARNING
Do not use P2-E6 in the presence of flammable anesthetics, detergents, or other chemical
vapors.
CAUTION
Do not block the air intake or air exhaust when operating the equipment. Blocked air
circulation or proximity to the heat source can cause internal heat build-up, shut down, or
damage to the concentrator.
CAUTION
The P2-E6 Concentrator is designed for continuous use. It is useful to operate the product
frequently for optimal sieve bed life.
CAUTION
P2-E6 is shipped from factory with battery removed.
Ensure the Pre-Filter Cover is in place.
Air Exhaust
Air Intake
CAUTION
Do not operate the device without the Cotton Intake Filter and Pre-Filter Cover in place.
Operating the device without these in place can damage the device.
19