Maintenance and cleaning of P2-E6
Battery Care and Maintenance - Continued
Effect of Temperature on Battery Performance
To extend the run-time of your battery, the device should be used in temperatures
between 41°F and 95°F (5°C and 35°C). The number of cycles that the battery will last is
highly dependent upon the temperature at which the battery is charged.
CAUTION
We suggest that the room temperature should not exceed 75°F (24°C) when the battery is
being charged.
Battery Time Remaining Clock
The P2-E6 continuously displays the battery time remaining. This displayed time is only an
estimate and the actual time remaining may vary from this value.
CAUTION
Store battery in a cool dry place with a charge of 40-50%. BATTERIES SHOULD NOT
BE LEFT DORMANT FOR MORE THAN 90 DAYS AT A TIME.
CAUTION
If the device is not used for an extended period of time, please remove the battery from the
device.
26