Alerts-Continued
Alarm item
Sieve Bed Fail
Compressor Fail
Valve Check Fail
Cooling Fan Fail
Battery Depleted
System Cold
Battery Cold
System Hot
Battery Hot
Power Supply
Fail
Gas Supply Fail
Sys Startup Fail
Alarm condition
Sieve Bed failure
or invalid
Compressor failed
Valve switch failed
Cooling fan failed
RSOC
5%
Without adapter
System Temperature
<32°F/0°C
Battery Temperature
<32°F/0°C
System Temperature
>140°F/60°C
Battery Temperature
>149°F/65°C
System voltage
<10.5v
Flow or concentration
below normal after
injection
Concentration less
than 87%
continuously >15s
after system startup
System process
Auto-shut down
after 10 seconds
Auto-shut down
after 10 seconds
Auto-shut down
after 10 seconds
Auto-shut down
after 10 seconds
Auto-shut down
after 10 seconds
Auto-shut down
after 10 seconds
Auto-shut down
after 10 seconds
Auto-shut down
after 10 seconds
Auto-shut down
after 10 seconds
Auto-shut down
after 10 seconds
Auto-shut down
after 10 seconds
Auto-shut down
after 10 seconds
15
Display of screen
Sieve Bed Fail
Contact Provider
Compressor Fail
Contact Provider
Valve Check Fail
Contact Provider
Cooling Fan Fail
Contact Provider
Battery Depleted
Replace battery/
Connect to adapter
System Cold
Shut down, Move to
warmer place
Battery Cold
Shut down, Move to
warmer place
System Hot
Shut down, Move to
cooler place
Battery Hot
Shut down, Only use
adapter
Power Supply Fail
Contact Provider
Gas Supply Fail
Contact Provider
System Startup Fail
Contact Provider