General Operation-Continued
7. Set flow rate prescribed by your provider.
Press the + or - setting buttons to adjust the P2-E6 to the desired flow rate. The current
setting can be viewed on the display from 1 to 6.
CAUTION
Ensure the power is in a well-ventilated place. During operation, the power supply may get
hot. Ensure the power supply is cool before handling.
8. Wear the nasal cannula on your face and breathe through your nose.
WARNING
If you feel discomfort using the device, consult your doctor immediately.
CAUTION
If the oxygen level drops below recommended levels, a Low O2: <87% alert will display on
the screen. If the alarm persists, contact your provider.
CAUTION
If there is no operation for 30 seconds, the display screen will dim. You can press any
button to light up the display.
CAUTION
The P2-E6 will notify you with an audible alert (if enabled) and a display showing "Absence
of Breath" if no breath has been detected for 15 seconds. After 15 seconds, the device will
enter auto-pulse mode until breath is detected. Once breath is detected, the device will
resume normal delivery of oxygen.
Operating Instructions
The P2-E6 will sense if you are breathing from it. If you are not
yet breathing through the cannula, the P2-E6 will begin to pulse
automatically about every 3 seconds.
As soon as you begin breathing through the cannula, the device will
begin delivering pulses based on your breathing. As your breathing
rate changes the P2-E6 will sense these changes and adjust the
amount of oxygen at your next inhale.
22