Maintenance and cleaning of P2-E6

Cleaning the Case

The outside case should be cleaned using a damp cloth with a solution of mild detergent
and water.
CAUTION
Do not allow liquids into any of the controls, the interior or the case, or the oxygen tubing
connector. If this occurs, contact your provider for assistance.
WARNING
Do not use alcohol, isopropyl alcohol, ethylene chloride or petroleum-based cleaners on
the cases or on the pre-filter coveres.

Cannula Replacement

The nasal cannula is disposable. You can buy replacements from your provider and follow
the cannula's manufacturer's instructions.
CAUTION
Nasal cannula should be FDA approved and the length should be less than 25ft (7.6m).

Filter Cleaning and Replacement

Filters are designed for adequate air flow through the device at the front of the P2-E6.
Pre-Filter Cover # P2E6PFCR
The Pre-Filter Cover must be cleaned weekly to ensure adequate air flow. Clean with a
mild liquid detergent and water. Ensure cover is completely dry before use.
CAUTION
It may be necessary to clean the Pre-Filter Cover more often in dusty or polluted
environments/conditions.