b) Under Standby interface, press Main button, and then the motor handpiece
will enter Working interface. The interface displays is as follows:
c) Press the Main button again, and then the motor handpiece returns to
Standby interface.
d) Hold down the Setting button "P", then press Main button to turn off motor
handpiece. In Standby Interface, the motor handpiece would automatically
shut down after 3 minutes without any button-pressing operation. The motor
handpiece will also automatically shut down when put on charge.
4.2 Selecting customized program sequence number
The motor handpiece has 10 memory programs(M0-M9) and 15 preset
programs, press Adjusting button "+"/"-" to change customized program
sequence number during standby state.
M0-M9 is a memory program for canal shaping and measurement, every
memory program has its own parameters such as Operation mode, speed and
torque, all these parameters can be changed.
4.3 Parameter setting
The below chart shows which parameters are able to be set in a specific
operation mode.
User Program Operation
Parameter
Set Working Speed
Set Torque Limit
Set Apical Action
Set Auto Start
Set Auto Stop
Set Flash Bar Position
Set Apical Slow Down
Set Forward Angle
Standby interface
Working interface
Mode
CW
YES
YES
YES
YES
YES
YES
YES
n/a
CCW
SGP
YES
YES
NO
YES
NO
YES
NO
YES
NO
YES
YES
YES
YES
NO
n/a
YES
14
CW-ATR
CCW-ATR
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
NO
NO
YES
YES
EAL
n/a
n/a
n/a
n/a
n/a
YES
n/a
n/a