3. Wet the dry soft cloth with 75% alcohol, and then wipe all surfaces of the
handpiece, charger, base and other components for 3 minutes.
4. Put the handpiece, charger, base and other components back into the clean
storage area.
Note:
a) The cleaning and disinfection must be performed within 10min before use.
b) The disinfectant used must be used immediately, no foaming is allowed.
c) In addition to 75% alcohol, you can use non-residue disinfectants such
as Cavicide, but you must respect the concentration,temperature and time
specifi ed by the disinfectant manufacturer.
6.4 The cleaning, disinfection and sterilization of contra-angle,lip
hook~probe areas
Unless otherwise stated, they will be hereinafter referred to as "products".
Warnings:
The use of strong detergent and disinfectant (alkaline pH>9 or acid pH <5)
will reduce the life span of products. And in such cases, the manufacturer
takes no responsibility.
The products may not be exposed to temperature above 138℃.
Processing limit
The products have been designed for a large number of sterilization cycles.
The materials used in manufacture were selected accordingly. However with
every renewed preparation for use, thermal and chemical stresses will result
in ageing of the products. The maximum number of sterilizations for products
is 250 times.
6.4.1 Initial processing
6.4.1.1 Processing principles
It is only possible to carry out effective sterilization after the completion
of effective cleaning and disinfection. Please ensure that, as part of your
responsibility for the sterility of products during use, only sufficiently
validated equipment and product-specifi c procedures are used for cleaning/
disinfection and sterilization, and that the validated parameters are adhered to
during every cycle.
Please also observe the applicable legal requirements in your country as well
as the hygiene regulations of the hospital or clinic, especially with regard to
the additional requirements for the inactivation of prions.
6.4.1.2 Post-operative treatment
The post-operative treatment must be carried out immediately, no later than
30 minutes after the completion of the operation. The steps are as follows:
1. Remove the products from the base, and rinse away the dirt on the surface
of handpiece with pure water (or distilled water/deionized water);
27