Pipetting
The volume is factory-set to the respective nominal volume of the ErgoOne® E Multi-Channel
pipette and can be easily and quickly changed individually, see Set volume.
NOTICE!
Perfect analysis results can only be achieved by using quality tips. We recommend the TipOne®
tips which are delivered with the pipette or other compatible pipette tips, see TipFinder on
Starlab's website www.starlabgroup.com.
If other pipette tips are used, check whether these tips fit the ErgoOne® E Multi-Channel
pipette and have the required quality for the desired application.
12 · ErgoOne
E Multi-Channel Pipettes · User Manual
®
a.
b.
c.
1x
Volume range
0.5 – 100 µl
100 – 300 µl
> 1000 µl
Insert tip vertically:
Use the correct tips, in accordance with the volume
range or color code!
Make sure that the tips are firmly in place and leak
tight. Pipette tips are disposable products!
Align pipetting unit: The pipetting unit can turn
freely in both directions.
Aspirate liquid: Hold the instrument vertically and
immerse the tip in the liquid. The liquid is aspirated
by pressing the pipetting key. The arrow in the
display points upwards (aspiration).
Leave the tip immersed in the liquid for a few
seconds, so that the set volume is aspirated
completely. This is especially important when
pipetting viscous media and when using pipettes
with large volumes.
Immersion depth
2 – 3 mm
2 – 4 mm
3 – 6 mm
Wait time
1 s
1 s
3 s