NOTICE!
ISO 8655 requires that pipette tips are pre-wetted once before the actual pipetting procedure.
4.5
Direct trigger blow-out
The blow-out can also be triggered directly at any time if necessary.
NOTICE!
During blow-out, the piston moves completely downwards. Ensure that any residual liquid is
discharged safely. Keeping the pipetting key pressed keeps the piston down and thus prevents
accidental aspiration of liquid. Releasing causes the piston to return to the start position.
d.
Dispense liquid: When the liquid aspiration is
finished, the arrow in the display points downwards
(dispensing). Place the pipette tip against the
vessel wall. Hold the pipette at an angle of 30–45°.
By pressing the pipetting key again, the liquid is
1x
dispensed completely with automatic blow-out
While doing this, wipe the pipette tip against the
vessel wall.
e.
Ejecting a tip: Hold the pipette Pipetting unit over
a suitable disposal container and the press the tip
ejection key.
a.
Call up the blow-out function: Press 'Enter' key.
1x
The display shows 'blo' for blow-out.
b.
Trigger blow-out: Press the pipetting key once
to trigger the blow-out. The display will then
jump back to the set pipetting mode (start
position).
1x
ErgoOne
E Multi-Channel Pipettes · User Manual · 13
®