b.
set revPIP mode: Use one of the arrow keys to
scroll through the modes until 'revPIP' appears.
'Mode' continues to flash.
1x
c. Confirm revPIP mode: Press 'Enter' key.
The display now shows 'blo' for blow-out.
d.
Preparing for pipetting: Press the pipetting key
once to move the piston to its start position.
The arrow in the display points upwards
(aspiration).
1x
e.
Aspirate liquid: Press the pipetting key once
to aspirate the liquid.
1x
NOTICE!
When aspirating the liquid, slightly more volume is
aspirated than set.
f.
Dispense liquid in revPIP mode: Press the pipetting
key once to dispense. On the display, the
arrow points down (dispensing). The set
volume is now dispensed and some liquid
1x
remains in the tip.
g.
Re-aspirate liquid in revPIP mode: Pressing
the pipetting key again now resumes the set
volume. (Pressing the pipetting key again
releases the volume)
1x
1x
h.
Trigger blow-out: After the last pipetting, press
'Enter' key.
The display now once again shows 'blo' for blow-
out.
ErgoOne
®
E Multi-Channel Pipettes · User Manual · 17