Integra MAYFIELD A1051 Manual De Instrucciones página 9

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 44
Phase
Time
Water
(Min.)
Temperature
Pre-wash
4:00
Cold water
1
Enzyme
4:00
Hot water
Wash
Wash 1
10:00
60.0ºC
(140º F)
Rinse 1
0:30
Hot water
Thermal
2:00
82.2ºC
Rinse**
(180º F)
** Optional phase for disinfection of components
– minimum water temperature as indicated or per
worker manufacturer specifications for the thermal
rinse cycle.
NOTE: Any deviation from this guideline could
result in damage to the equipment as well as
improper cleaning results.
Rinse with purified water. Do not perform if
parameters cannot be achieved.
5. Remove from washer and dry completely if
needed.
6. Inspect equipment to make sure there is no
visible organic debris or residue from cleaning
agent.
Repeat process if any visible soil is detected.
The life expectancy of the MAYFIELD products is
expected to be as long as 7 years.
Maintenance and Care
To ensure proper function and to extend the life and
performance of the equipment, Integra LifeSciences
recommends the following:
Detergent and
Recommended Action
Concentration
Return the device to
N/A
the Integra LifeSciences
Repairs department for
Neutral pH
detailed inspection and
enzymatic (e.g.
servicing.
Endozime® AW
Request that Integra
Triple Plus with
NeuroSpecialists perform
APA, Ratio 1:128)
routine inspections of the
Neutral pH
device
detergent
(e.g. Renu-
In the absence of proper care and servicing of the
Klenz™ , Steris,
device, negative effects may be seen after repeated
Ratio 1:256)
processing over time which may lead to reduced
N/A
performance.
N/A
Contact information: See the Service and Repair
section for contact information on how to return
your device for periodic servicing and to request
periodic inspections.
See Inspection and/or Service notes section for
routine checks to be performed on the device.
NOTE: Any serious incident that has occurred in
relation to the device for the user and/or the patient
should be reported to the manufacturer and the
competent authority of the member state in which
the user and/or patient is established.
Device Disposal
NOTE: Follow hospital procedures for disposal of
this device.
9
EN – ENGLISH
Recommended Frequency
Once / year
Twice / year