8 Cleaning and maintenance work for the user
To ensure long product life and scan accuracy, it is necessary to regularly clean and calibrate the
product.
8.1
Recalibration
– The calibration plate must not be cleaned. Touching the reference points located on the
calibration plate can cause the calibration accuracy to decrease.
– Make sure that the reference points of the calibration plate are not damaged as otherwise
calibration is no longer possible.
– In case of a malfunction that cannot be resolved by restarting the scanner or workstation,
contact your customer support.
To guarantee consistent scan results, we recommend you calibrate the scanner on a weekly basis
using the supplied calibration plate.
The scanner should also be calibrated after transportation or if the scanner is moved. Instructions
for recalibrating your scanner can be found in our PrograScan software manual. Initiate the scan
process using the software and follow the instructions step by step. If you are unsure, please
contact our customer support.
8.2 Cleaning
DANGER! Accidentally switching on the scanner during cleaning.
Death or severe personal injury due to electric shock.
– Before cleaning, take the optical 3D scanner out of operation and safeguard it from being
started without authorization.
– Turn the scanner off using the switch on the rear panel to eliminate any risk of injury caused by
accidentally switching on the scanner.
NOTICE! Scanner exposed to water, steam or cleaning agent.
Damage to the electronical parts of the scanner.
– Ensure that no water, steam or cleaning agents penetrate the electrical scanner parts during cleaning.
NOTICE! Improper cleaning of the optical sensor. The optical sensor located in the upper part of
the scanner is is extremely sensitive.
Damage to the optical sensor.
– Do not clean the optical sensor system located in the upper part of the scanner.
– Do not use harsh cleaning agents, as they can also damage the scanner.
– Ensure that the scanner is only cleaned by people who are trained and who have been informed
of the scanner's potential dangers.
– Stop the scanner and disconnect it from the power supply for the duration of the cleaning.
– Vacuum or wipe the scanner with a dry cloth.
– Pay attention to the optical components when cleaning. These must not be adjusted.
– To prevent damage to the optical sensors located in the upper part of the scanner, they may only
be dusted with a microfibre cloth.