EMC for using this product in the medical environment
• The description of GP-UH532 represents the camera system using GP-UH532C and GP-UH532H or GP-UH332H.
WARNING
• Usage of accessories or cables that are not specified in this document may increase electromagnetic radiation to be emit-
ted from GP-UH532 or deteriorate the immunity to the electromagnetic interference.
• If using GP-UH532 by installing next to or stacking on other device, make sure that the CCU works normally in the system.
CAUTION
• To dispose of this product, it is required to comply with the local or national legislation and related hospital policies.
Policy and declaration by manufacturer - Electromagnetic emission
GP-UH532 is intended to be used in the electromagnetic environment specified below.
Ensure that this product is used in the following environment.
Emission test
Conductive emission of radio
frequency
CSPR 11
Conductive emission of radio
frequency
CSPR 11
Harmonic current emissions
IEC 61000-3-2
Voltage variation, voltage
fluctuations and flicker emis-
sion
IEC 61000-3-3
Use the following AC Adapter:
Manufacturer: Sinpro Electronics Co., Ltd.
Part No.: MPU60A-105-P49L10 *
Compatible level
Group 1
GP-UH532 uses radio frequency energy for internal functions. Therefore,
radio frequency emission is considerably low, and it will not interfere with
electronic devices placed nearby.
Class A
Due to the emission characteristics of GP-UH532, it is suitable for use in
the industrial area and hospitals. If this product is to be used in the resi-
dential district, it may require taking some measures, relocation of the
product for example, since the protection for the radio frequency com-
munication device may not be sufficiently provided in the residential dis-
Class A
trict.
Clause 5
Electromagnetic environment - Policy
3