Operating instructions
3. Description
GB
2. Function: selects the fi le systems in use. Allows reloading the unit's default
parameters (by simultaneously pressing Function
3. Files: selects the fi le being used. Each fi le is associated with the correct torque
and speed values.
4. Torque: modifi es the torque of the fi le being used.
5. Rpm: modifi es the rotation speed of the fi le being used.
6. Reverse: (LED on) sets a counter-clockwise rotation direction for the fi le. By holding
down the button, saves the fi le parameters on the memory stick provided.
7. A/Rev: (LED on) sets the automatic reverse for the micromotor. The fi le changes
rotation direction as soon as the torque set is reached. By holding down the
button, reloads the fi le parameters from the memory stick provided.
8. Cal: calibrates the contra-angle to ensure torque accuracy each time the contra-
angle is replaced or lubricated.
9. Anatomy: reduces the torque with respect to the torque set, depending on the
diffi culty of the endodontic treatment (root canal morphology).
10. Red: allows choosing between several reduction ratios of the contra-angle being
used.
11. Motor: starts the motor from the keyboard rather than by using the pedal. The
motor starts if it is pressed once. Pressing it again stops the motor.
Foot Pedal Control
ATR EndoPocket's micromotor can also be started up using the pedal. Contrary to the
"Motor" button on the keyboard, the micromotor operates as long as the switch remains
pressed. Pushing down on the pedal once after the motor has been started using the
keyboard stops the motor (the pedal overrides the button).
Advanced Technology Research s.r.l.
and Function