6. Use with EP-Fill (sold as an accessory)
ATR EndoPocket is able to drive the ATR EP-Fill device. ATR EP-Fill is a handpiece that
allows vertical obturation to be performed. Please consult the ATR EP-Fill instruction
manual for any information concerning the device.
✓ Connect ATR EP-Fill and the pedal to ATR EndoPocket. Use the micromotor
connector to connect ATR EP-Fill. Ensure the connection is well inserted.
✓ Once the ATR EP-Fill connector is connected to ATR EndoPocket, the following
screen is displayed:
✓ Insert the plugger desired (see paragraph 4 Description of the ATR EP-Fill
instruction manual).
The unit can be set to 3 different temperatures: 200°C, 250°C and 300°C, which can be
selected using the Torque
ATR EP-Fill uses the pedal provided with ATR EndoPocket to activate the dental
handpiece heating function. Pressing the pedal activates the heating function. Keeping
the pedal pressed down allows ATR EndoPocket to reach the temperature set on the
plugger and will keep it steady over time. During the period in which the temperatures
increases, the LED on the MOTOR button remains lit. Once the temperature set has been
reached, ATR EndoPocket will emit a short acoustic signal. While the temperature is
being kept steady over time, the LED on the MOTOR button will fl ash and be accompanied
by a continuous acoustic signal.
Keep the pedal pressed down to complete the procedure.
If the pedal is released for a few seconds and then pressed down again, ATR EndoPocket
will begin heating the plugger again starting at its current temperature; therefore the time
needed to increase the temperature (MOTOR LED constantly lit) in this case can be quite
brief or none at all.
button.
19
e
P
ndo
ocket
Operating instructions