13. Accompanying Documents
ATR EndoPocket is intended for use in the electromagnetic specifi ed below. The customer
or the user of the ATR EndoPocket should only be used in such an environment.
Test level
Immunity test
IEC 60601
Conducted RF
3Vrms
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3V/m
IEC 61000-4-3
80MHz to
2.5GHz
Compliance
Electromagnetic environment - guidance
level
Portable and mobile RF communication
equipment should be used no closer to any
part of the ATR EndoPocket, including cables,
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter
3 V
Recommended separation distance
d = [3.5/V
3 V/m
d = [3.5/E
d = [7/E
where P is the maximum output power rating
of the transmitter in watts (w) according to the
transmitter manufacturer and recommended
separation distance in metres (m).
Field strengths from fi xed RF transmitter, as
determined by an electromagnetic site survey
a
should be less than the compliance level in
each frequency range
Interference may occur in the proximity of
equipment marked with the following symbol:
29
e
ndo
Operating instructions
]√P
1
]√P 80MHz to 800MHz
1
]√P 800MHz to 2.5GHz
1
b
.
P
ocket
GB

Publicidad