Recommended Separation Distances - NUK 10.256.345 Operación Manual

Ocultar thumbs Ver también para 10.256.345:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Immunity test
Power
3 A/m
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Note: UT is the a.c. mains voltage prior to application of the test level.
between portable and mobile RF communications equipment and the ME equipment or
ME system.
The THixyz series is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the THixyz series can help
prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the THixyz series as
IEC 60601 test level

9.3 Recommended separation distances

Compliance
Electromagnetic environ-
level
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
- 55 -
ment – guidance