3. Slide your hand along a deflated cell under the patients
sacral area (bottom). The inner static cell will remain
inflated but your hand should slide easily between
patient and base.
4. If hand can pass under patient then patient is adequately
suspended and pressure can be lowered.
5. Repeat Bottoming Out test after pressure has been
lowered.
In the event of a system malfunction, the alarm will
activate and pressure LEDs will flash.
Establishing Pressure (inclined position)
The system runs on the chosen setting but when
the user changes and stays in a new position the
pressure in the mattress can adjust (higher/lower).
The sensor in the control unit will react and adjust
the pressure if required to compensate for the new
position.
IMPORTANT!
– Return to the original pressure setting when
the patient returns to the supine position, and
perform a Bottoming Out test.
– Wait a minimum of 10~12 minutes between
pressure adjustment and patient assessment,
as it may take a full cycle for the system to
adjust to any new setting.
5.4 Heel cell zones - Softcloud Pro
For Softcloud Pro, the 5 heel cells incorporate an option
which allows the carer to permanently offload the heel of
1630732-A
the user, ideal for the most vulnerable users such as those
with limb ischaemia or existing wounds.
CAUTION!
Insufficient pressure relief
– Deflating more than one adjacent air cell can
affect the support of the user during the
normal alternating support cycle. This function
should only be used for temporary procedures.
A
B
1. Open the push button flap A at the foot end of the
mattress to access the 5 release connectors.
2. Disconnect the chosen connector B for permanent
offloading.
•
Top connector is closest to the knee.
•
Connector 4–2 is below the knee and goes
towards the heel.
•
Bottom connector is the last heel cell
closest to the end of the mattress.
It is not recommended to deflate more than one
air cell directly beneath the area where relief
is needed.
Usage
19