THERAPY CYCLES, PROFESSIONAL ONLY
1. Turn the unit on,
set the temperature setpoint.
30) minute cycle, LED should be solid, and the selected temperature LED is flashing.
2. Once the T/Pump reaches the desired temperature, the Warming LED (Red) flashes with a short audible beep.
This is to let the operator/user know the unit is at temperature and the timed therapy is starting.
3. Once the 20 (or 30) minute period is up, the heater and pump shuts off. This stops any flow to the pad. The
Setpoint and Therapy time LED's will be blinking during the off Time period.
4. Once the Off Time period is up, the pump and heater will restart. Time Therapy LED goes solid, and the
Setpoint LED is flashing. Once the Setpoint is reached the Red LED flashes with a short audible beep to let the
operator/user know the next therapy cycle has started.