2.8 Initial Setup (Cont.)
2.8.4 Setup Menu (Cont.)
Press G/N/T/ MENU , , , , , F
scale capacity selections. xx= value last set.
Selections are:
5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120,
150, 200, 250, 300, 400, 500, 600, 750, 1000,
1200, 1500, 2000, 2500, 3000, 5000, 7500,
10000, 20000 (lb or kg).
Press PRINT/ UNITS button until desired capacity
value is reached.
Press G/N/T/ MENU b utton, Gd0.01 is displayed.
This is the graduation size. Refer to paragraph
2.8.3 Load Cell Capacity Information table. For
available selections, press PRINT/ UNITS button until
desired graduation value is reached.
Press G/N/T/ MENU button, CP 30 kg is displayed.
This is the full scale calibration point setting. The
range is from 20% to 100% Full scale capacity.
Press PRINT/ UNITS button until desired calibration
value is reached.
Press G/N/T/ MENU button, AS 3 is displayed. This
is the duration of animal sample time which can be
set from 1 to 10 seconds.
Press PRINT/ UNITS button until desired sample time
value is reached. See Note below.
Press G/N/T/ MENU button, Ad 4 is displayed. This
is the duration of animal weighing display time
which can be set from 1 to 10 seconds. The default
is 10 seconds. See Note below.
Press PRINT/ UNITS button until desired sample time
value is reached. See Note below.
Press G/N/T/ MENU button to end this block, SAVE is
displayed.
Press G/N/T/ MENU button, rEAD is displayed which
is the next menu or press PRINT/ UNITS button to
return to Setup menu.
The Indicator is now matched up with the scale
base and the Indicator parameters may now be set
and calibrated.
NOTE: Not available if LFT is ON. Only available if ALT
is set to ANI.
CW-11 Indicator
xx is displayed. . . . . This is full