2.8 Initial Scale Setup (Cont.)
2.8.7 Lockout Switch Menu (Cont.)
Center
Gross
Brutto
of Zero
Procedure (Cont.)
With the Indicator ON, press and hold the G/N/T/
MENU button until MENU is displayed. When you
release the G/N/T/ MENU button, CAL is displayed,
then press PRINT/ UNITS button, until LOCSW is
displayed.
Press G/N/T/ MENU button, LSTOFF is displayed.
This permits locking the Setup menu. OFF is un-
locked, ON is read only (locked). This menu is
hidden if the CAL jumper is off.
Press PRINT/ UNITS button, and select ON or OFF.
Press G/N/T/ MENU button, LrdOFF displayed. This
permits locking the Readout menu. OFF is un-
locked, ON is read only (locked).
Press PRINT/ UNITS button, and select ON or OFF.
Press G/N/T/ MENU button, LPtOFF is displayed.
This permits locking the Print menu. OFF is un-
locked, ON is read only (locked).
Press PRINT/ UNITS button, and select ON or OFF.
Press G/N/T/ MENU button, LCLOFF is displayed.
This permits locking the Calibration menu. OFF is
unlocked, ON is read only (locked). This menu is
hidden if the CAL jumper is off.
Press PRINT/ UNITS button, and select ON or OFF.
Press G/N/T/ MENU button to end this block, SAVE
is displayed.
Press G/N/T/ MENU , Quit is displayed..
Press PRINT/ UNITS button to go to CAL or press
G/N/T/ MENU button, Indicator returns to a weighing
mode.
NOTE: At this point, the Indicator must be calibrated
and the jumper removed from the CAL connector in
order to lock out the menus. The top cover of the
Indicator should be free to gain access to the CAL
jumper.