EN-24
GEOGRAPHICAL FACTOR (For Europe Only)
Press the G/N/T/ Menu button, gEO 19 is displayed. This is the current geographi-
cal adjustment value.
The geo factor includes settings from 0 to 31 and is used to compensate for
varying gravity at different geographical areas (complete geographical adjustment
settings are listed in the following table).
Only an authorized manufacturer's representative or certified verification
personnel may make these changes.
Changing the geographical setting alters the calibration values.
Press the Print/ Units button until the desired geographical adjustment value is
reached. The factory default setting is gEO 19
SAVE
Press the G/N/T/ Menu button to end this block, SAVE is displayed.
Press the G/N/T/Menu button to save the geographical factor setting. The next
menu SEt uP is displayed.
or
Press the Print/ Units button to go back to the CAL menu without saving changes to
the geographical setting.
QUIT
To exit from the CAL menu, press the Print/ Units button to advance to Quit. Then
press the G/N/T/ Menu button to go back to the weighing mode.
CW-11 Indicator
NOTE: