2.8 Initial Setup (Cont.)
2.8.5 Readout Menu (Cont.)
Procedure
Procedure
Procedure
Procedure
Procedure
AVERAGING LEVEL (Cont.)
Press G/N/T/ MENU button, AL LO is displayed. This is
averaging level settings. Selections are:
'Lo' - Averaging level is low
'Hi' - Averaging level is high.
Press PRINT/ UNITS button,and select LO or HI.
STABILITY
The stability range specifies the weighing results and
must be within a preset tolerance limit for a certain
time to turn the stability indicator ON. When a dis-
played weight changes beyond the allowable range,
the stability indicator turns OFF, indicating an unstable
condition. Factory default setting is shown in bold
type.
.5d Smallest range: stability indicator is
ON only when displayed weight is
within .5 divisions.
1d Normal setting. - Fixed for LFT
3d More stable course
Press G/N/T/ MENU button, StAb1 is displayed.The
stability range specifies the weighing results and
must be within a preset tolerance limit for a certain
time to turn the stability indicator ON. When a
displayed weight changes beyond the allowable
range, the stability indicator turns OFF, indicating an
unstable condition. 0.5 d smallest range, stability
indicator is ON only when displayed weight is stable
within 0.5 divisions. 1 d-stable within 1 division. 3
d-largest range, stability indicator is ON even
though displayed weight changes 3 divisions.
Factory default setting is 1.
Press PRINT/ UNITS button,and select 0.5, 1 1 1 1 1 , or 3.
Normal 1d stability is default/recommended.