6607290.book Seite 46 Donnerstag, 4. Januar 2018 11:55 11
7 Conditioning
46
Operating Instructions Midwest Stylus
1. Brush the high-speed handpiece under running water (< 38 °C,
< 100 °F, at least drinking water quality) and good lighting (min. 500
lux) and color rendering index (min. 80 Ra) until no more dirt can be
seen, for at least 10 seconds.
2. Flush the drive channels with spray.
3. Maintain mechanical parts manually [ → 47].
4. Maintain the push button chuck manually [ → 50].
5. Pack the high-speed handpiece in packaging material suitable for
sterilization and storage. e.g. paper/plastic composite packaging.
6. Conduct sterilization [ → 54].
®
®
Plus / Midwest Tradition
TC/Pro TCF
66 07 290 D3673.201.01.04.02